Abstract: 

Between 1994 and 1999, 217 metal-on-metal total hip arthroplasties with a low-carbide bearing surface were performed with use of the cementless Zweymüller SL-Plus stem and the Bicon-Plus threaded cup in 194 consecutive patients. After a minimum follow-up of ten years, 181 living patients (203 hips) were available for evaluation. The revision rate after an average of twelve years was 18% (thirty-six hips in thirty-six patients were revised). The main reason for revision was aseptic loosening of one or both components. The probability of survival of the stem at fifteen years was 77% (95% confidence interval [CI], 65% to 86%). The probability of survival of the cup was 80% (95% CI, 62% to 90%). These high failure rates at mid-term follow-up led us to abandon the use of low-carbide metal-on-metal total hip arthroplasty components.

Level of Evidence: 

Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.