Effect of a Cast on Short-Term Reproducibility and Bone Parameters Obtained from HR-pQCT Measurements at the Distal End of the Radius

首页 >> 文献期刊 >> 正文

Effect of a Cast on Short-Term Reproducibility and Bone Parameters Obtained from HR-pQCT Measurements at the Distal End of the Radius

第一作者:Joost J.A. de Jong 编号 : #127162#
2016-03-11
我要评论

Joost J.A. de Jong, Jacobus J. Arts, Ursina Meyer, Paul C. Willems, Piet P. Geusens,  

Joop P.W. van den Bergh,  Bert van Rietbergen,


Background:

High-resolution peripheral quantitative computed tomography (HR-pQCT) is a promising tool to assess the fracture-healing process at the microscale in vivo. Since casts are often used during fracture treatment, they might affect the assessment of bone density, microarchitectural, and biomechanical parameters and the short-term reproducibility of those parameters, e.g., as a result of beam-hardening. The aim of this study was to assess the effect of a plaster-of-Paris and/or fiberglass cast on bone parameters and on the short-term reproducibility of the HR-pQCT measurements of those parameters.

Methods:

The effects of a cast on HR-pQCT-derived bone parameters were evaluated by comparing HR-pQCT scans of fifteen human cadaveric distal radial specimens from one male and fourteen female donors (median age, eighty-four years [range, sixty-two to ninety years] at the time of death) in three conditions: with a plaster-of-Paris cast, with a fiberglass cast, or without a cast. Short-term reproducibility was assessed using duplicate scans of the distal end of the radius in sixteen healthy volunteers without a fracture (nine men and seven women with a median age of twenty-six years; range, twenty-two to thirty-nine years) while wearing and not wearing a fiberglass cast.

Results:

Compared with measurements made with no cast, the plaster-of-Paris cast introduced a systematic error in the bone parameters ranging from −2.6% in trabecular separation to −9.8% in cortical thickness. Bone parameters were affected only marginally by fiberglass, with errors between −0.6% and −1.6% in trabecular separation and cortical thickness, respectively. Short-term reproducibility with a fiberglass cast was similar to that with no cast: approximately 1% for bone density parameters, 4% to 5% for microarchitectural parameters, and 3% to 4% for biomechanical parameters.

Conclusions:

A plaster-of-Paris cast has a considerable effect on HR-pQCT measurements. A fiberglass cast only marginally affects the bone parameters, and the short-term reproducibility of HR-pQCT measurements in patients with a fiberglass cast is comparable with that in patients without a cast. In studies on fracture-healing using HR-pQCT, a fiberglass cast is desirable if immobilization is indicated. The use of a plaster-of-Paris cast should be avoided if possible; however, if not avoidable, corrections after the scan are desirable to adjust for the error introduced in the bone parameters.